- Want to use external raster control for 20-066 ion gun control to allow SE imaging for focusing and alignment
- Omicron software does not allow external raster to be enabled
- Install PHI software (from floppy disk!!) version 3.0A 1997
- C:\program files\PHI\EXE
- edit IonGunCrtl.ini
[Ion Gun Control]
ComPort=COM6
RS485ResponseTime=1000
IgunAddress=3
NoHardware=0
ModelNumber=11-066
BlankingOptics=1
[END]
Software seems to work.
Pressure control
- Phi software does not generate setpoint voltage
- switch RVG50C Pfeiffer controller to INT setpoint
- knob 1.0 = 13 mPa
- 1.5 = 20 mPa
- 2.0 = 26.5 mPa
- Leak valve needs 2.0 turns for ~ 30 mPa (as of 12/2/13)
PHI software raster
raster is not centered on offset, ie changing raster size shifts sputter area
Y raster half of X gives approximate square image for 30 degree tilt sample