Friday, December 6, 2013

Ion gun imaging


  • Want to use external raster control for 20-066 ion gun control to allow SE imaging for focusing and alignment
  • Omicron software does not allow external raster to be enabled
  • Install PHI software (from floppy disk!!) version 3.0A 1997
    • C:\program files\PHI\EXE
    • edit IonGunCrtl.ini

[Ion Gun Control]
ComPort=COM6
RS485ResponseTime=1000
IgunAddress=3
NoHardware=0
ModelNumber=11-066
BlankingOptics=1
[END]
    Com port set to COM6, address used by Omicron. Other params left at defaults

    Software seems to work.

    Pressure control

    • Phi software does not generate setpoint voltage
    • switch RVG50C Pfeiffer controller to INT setpoint
      • knob 1.0 = 13 mPa
      • 1.5 = 20 mPa
      • 2.0 = 26.5 mPa
    • Leak valve needs 2.0 turns for ~ 30 mPa (as of 12/2/13)

    PHI software raster

    raster is not centered on offset, ie changing raster size shifts sputter area
    Y raster half of X gives approximate square image for 30 degree tilt sample