Prior status
- just after bake, EHT off current up 246 vs 232
- Gun align near user software limit, stig balance off, time for full align
Align
8 kV aperture image condenser at "max" |
aperture positions:
- 60 um center
- 20 um right
- 30 um lower left
- 75 um lower right
- 90 um left
- 120 um top right
- blank top left
Procedure
First check mechanical alignment, use Si 10 um grid for test
Alignment, all offsets zero user and service. Cond normal
condenser off
Clear problems with mechanical alignment
Bring aperture to exact center, bad illumination
very poor aperture illumination on exact mechanical center
condenser on, illumination distorted here too
Procedure says illumination trumps centering
illumination better but is 120 better than 60??
best mechanical source alignment
after this use gun offset/align for center aperture